Locating atoms with higher precision

Two major problems that limit the resolution and interpretation of electron microscopy images are lens aberrations and multiple scattering. Chen et al. overcame these issues with ptychography, a technique that uses coherent scattering and multiple overlapping illumination spots to reconstruct an image from far-field diffraction patterns. This method works at a resolution that is limited, not by optics, but rather by the scattering strength of the sample, so it can work better with thicker samples. The authors achieved ultimate lateral resolution better than the thermal vibration of atoms in a PrScO3 sample and showed that it is theoretically possible to identify single dopant atoms.
Science, abg2533, this issue p. 826

Abstract

Transmission electron microscopes use electrons with wavelengths of a few picometers, potentially capable of imaging individual atoms in solids at a resolution ultimately set by the intrinsic size of an atom. However, owing to lens aberrations and multiple scattering of electrons in the sample, the image resolution is reduced by a factor of 3 to 10. By inversely solving the multiple scattering problem and overcoming the electron-probe aberrations using electron ptychography, we demonstrate an instrumental blurring of less than 20 picometers and a linear phase response in thick samples. The measured widths of atomic columns are limited by thermal fluctuations of the atoms. Our method is also capable of locating embedded atomic dopant atoms in all three dimensions with subnanometer precision from only a single projection measurement.
Transmission electron microscopy (TEM) plays critical roles in the study of micro- and nanostructures in many fields, including physics, chemistry, structural biology, and materials science. Progress in aberration-corrector optics for electron microscopes (1, 2) has substantially improved the quality of the imaging system, pushing the spatial resolution to the sub–50-pm level (3, 4). Nevertheless, in practical samples, this resolution limit is achievable only in extremely favorable conditions. One major obstacle is that multiple electron scattering is unavoidable in samples thicker than a monolayer, owing to the strong Coulomb interaction between the beam electrons and the electrostatic potentials from the atoms (5). Multiple scattering changes the beam shape within the sample and leads to a complicated intensity distribution at the detector plane. Nonlinear or even nonmonotonic contrast dependences on the sample thickness occur when imaging samples thicker than a few dozen atoms, which hampers direct structure determination by phase-contrast imaging. Quantitative structure-image interpretation usually relies on intensive image simulations and modeling (6). Direct retrieval of the sample potential requires solving the nonlinear, inverse problem of multiple scattering. Efforts have been made through different approaches, most of which are based on Bloch wave theory, by phasing different Bragg beams of crystalline samples (7, 8). Unfortunately, these approaches become extremely difficult to implement for general samples with large unit cells or aperiodic structures, because a large number of unknown structure factors needs to be determined.
Ptychography is another phase-retrieval approach that dates back to Hoppe’s work in the 1960s (9). Modern, robust setups use multiple intensity measurements—usually a series of diffraction patterns collected by scanning a small probe across the extended sample (10). No periodicity or symmetry constraints on the sample structure are required as a priori knowledge. This approach has been widely used in the visible-light imaging (11) and x-ray imaging (12) communities. Until recently, electron ptychography has been constrained by sample thickness and the limited detector performance in electron microscopy. Two-dimensional (2D) materials and the development of direct-electron detectors have led to a wider renewed interest (1315). Electron ptychography for thin samples, such as 2D materials, has achieved an imaging resolution of 2.5 times the diffraction limit of the lenses, down to a 39-pm Abbe resolution (15). However, such super-resolution approaches can be applied reliably only to samples thinner than a few nanometers, and the resolution in thicker samples is not substantially different from that attained with conventional methods (16). Such thin samples are practically difficult to achieve for many bulk materials, which currently restricts applications to 2D-like systems, such as twisted bilayers (17). For samples thicker than the probe’s depth of focus, multislice ptychography, using multiple slices to represent the sample, has been proposed (18). The structures of all slices can be retrieved separately. There are several successful demonstrations of multislice ptychography using either visible-light imaging (19) or x-ray imaging (20, 21). Owing to experimental challenges, however, only a few proof-of-principle multislice electron ptychography demonstrations (14, 22) have been reported and are limited in either resolution or stability.
We demonstrate multislice electron ptychographic reconstructions experimentally, recovering a linear phase response versus thickness, and push the lateral resolution close to the intrinsic atomic size, limited by thermal fluctuations of the atoms themselves. The experimental setup is shown schematically in Fig. 1A. A defocused electron probe is raster-scanned across a slablike sample, with one electron diffraction pattern recorded at each probe position by a high–dynamic range pixel array detector (fig. S1) (23). For thick samples, the probe function within the sample changes shape as a result of both wave propagation and strong dynamical scattering. A profile of the probe’s evolution with depth into the sample is shown in Fig. 1B. In conventional ptychography, we approximate the wave function at the exit surface of the sample as a multiplication of the incident wave function with a single projected sample function (10, 24). For thick samples, however, portions of the sample at different depth positions are effectively illuminated with different wave functions as a result of beam spreading by diffraction. Following the well-known Cowley-Moodie multislice solution of the electron dynamical scattering problem (5), the sample can be considered as many thin slices until each slice satisfies the multiplicative approximation (10, 12). The entire scattering procedure is modeled as sequential scattering from each slice followed by a free-space propagation to the next slice. For the inverse problem, a similar multislice procedure can be adopted on each iteration, as in multislice ptychography (18). In this work, there is no a priori assumption about the sample structure, and reconstructions start from random initial phases for all slices. The specimen potential for each slice is separately recovered through the phase of the transmission function, as illustrated in Fig. 1C (23).
Fig. 1 Principle of multislice electron ptychography.
(A) Experimental setup for electron ptychography. (B) Intensity distribution of electron probe during free-space propagation. The sample position is indicated by a dashed rectangle. (C) Transmission functions of multiple objects at different depth positions reconstructed via multislice electron ptychography, ψ i i = 1 , 2 , ... , N . (D) Total phase images, summed from each 0.5-nm-thick slice, reconstructed via multislice electron ptychography (MS-ptycho) using simulated diffraction data for PrScO3 [001] samples with different thicknesses. Corresponding phase images reconstructed by means of single-slice electron ptychography (SS-ptycho) are also shown for comparison. Scale bar, 2 Å. A structural model of PrScO3 is shown at right. (E) Phases at Pr, Sc, and O atomic sites [denoted by a circle, triangle, and rectangle, respectively, in the top-left image of (D)] from samples with different thicknesses obtained via multislice ptychographic reconstructions. Linear fits are shown as dashed lines.
We first compare the performance of multislice and single-slice electron ptychography on datasets simulated for crystalline PrScO3, the same system used for our experimental measurements. For samples as thin as 8 nm, our simulation shows that multislice electron ptychography gains resolution improvement (Fig. 1D) over the single-slice approximation. For thicker samples up to 30 nm, multislice electron ptychography demonstrates stronger performance improvements, showing the 59-pm separation of Pr–Pr dumbbells, whereas single-slice electron ptychography fails to capture even the basic structure. For even-thicker samples, such as those with 50-nm thickness, reconstructions are still possible, but higher sampling densities (21) are required to obtain a convergent solution from multislice electron ptychography (fig. S2).
Multislice electron ptychography provides quantitative phase information, with the phase increasing linearly as more layers are added into the sample. Figure 1E shows the phase change at different atomic positions from reconstructions with different sample thicknesses. Such linearity is crucial for retrieving 3D structural information, especially for phase-contrast electron tomography, which requires a monotonic contrast dependence of sample thickness (24). Nevertheless, conventional methods such as scanning TEM (STEM) annular dark-field and annular bright-field imaging or high-resolution TEM (HRTEM) imaging always have a nonlinear or even nonmonotonic dependence on thickness (figs. S3 and S4), and extensive simulations are required for quantitative image interpretation. Additionally, fast data acquisition and low-dose imaging capabilities that outperform conventional imaging techniques are retained in multislice ptychography, as in single-slice ptychography (figs. S10 and S11) (17). At present, a conventional imaging mode is still needed to align the microscope and survey the sample, as the present detector technologies and reconstruction algorithms are still too slow for live feedback.
We shift to experimental data collected on a 300-keV Titan Themis instrument with an electron microscope pixel array detector (EMPAD) (25). Full experimental details are given in the supplementary materials (23). Figure 2A shows one region of the phase image reconstructed by using an experimental dataset from a PrScO3 sample with a thickness of 21 nm projected along the [001] zone axis, where the c axis of PrScO3 is taken as the longest axis; this is the nonstandard Pbnm setting of space group 62 (a = 5.61 Å, b = 5.78 Å, and c = 8.03 Å) (26). The phase image resolves all atoms in the structure and outperforms state-of-the-art conventional electron microscopy (fig. S6) in terms of contrast and resolution: The heavy-metal atom sublattices for both Pr and Sc and the light-atom sublattice for O are resolved with a high contrast and signal-to-noise ratio. For these experimental data, both a partial coherence treatment of the electron probe (17, 27) and multiple slices (18, 21) are critical factors for obtaining a high-quality reconstruction (fig. S7).
Fig. 2 Multislice electron ptychographic reconstruction of PrScO3.
(A) Total phase image, from summing all slices, of [001]-oriented PrScO3 from a multislice electron ptychographic reconstruction of experimental data. Scale bar, 2 Å. (B and C) Phase profiles across Pr–Pr dumbbells and O–Sc–O directions, respectively. Red shading illustrates the variation from ~60 profiles at different atomic columns; black lines are their averages. Example positions are marked as blue lines in (A). (D) Intensity of Fourier transformation of the reconstructed phase image. Scale bar, 1 Å−1. (E) Power spectrum along the direction marked by the dashed arrow in (D). The intensity of the power spectrum is weighted as a power of 0.2 to highlight the high-frequency information transfer. The Bragg peak of 020 is labeled in (E).
The high–spatial resolution phase image of Fig. 2A is borne out by quantitative analysis. In real space, the Pr–Pr dumbbells with a separation of only 59 pm are resolved with a contrast of 63% (Fig. 2B), which is better than the 73% contrast for two point objects separated at the Rayleigh criterion. Therefore, the Rayleigh resolution of the image is much better than 59 pm. Nevertheless, the exact resolution can be determined only after considering the finite atomic size instead of assuming point objects (28). We can also resolve the O–Sc–O triple atom projections, even though the light O atoms are only 63 pm (26) from the heavier Sc atoms (Fig. 2C), and these cannot be resolved with conventional imaging techniques. The power spectrum from the Fourier transformation of the phase image (Fig. 2, D and E) shows an isotropic information transfer that is larger than 4.39 Å−1, corresponding to 23 pm in real space.
To estimate the spatial resolution of the ptychographic reconstruction, the intrinsic width of the atoms needs to be taken into account. The static projected potential of PrScO3 (i.e., at 0 K and neglecting zero-point thermal fluctuations) is very narrow (Fig. 3A). Nevertheless, the experiments were carried out at room temperature (300 K), and thermal fluctuations of atoms greatly broaden the potential (Fig. 3B). Additionally, imaging conditions, such as the finite illumination dose and maximum scattering angle of the collected diffraction patterns (15), will further impose a broadening factor in the reconstructed phase image (Fig. 3C). The combined effects of limited resolution and thermal fluctuations on the measured potential (Fig. 3D), each being roughly Gaussian in profile, can be added in quadrature and approximately modeled as a convolution of one Gaussian function with the static, frozen potential (23). From the experimental data, the measured widths [full widths at half maximum (FWHMs)] of each atomic column estimated from more than 60 atomic columns are 44 ± 1 pm, 45 ± 1 pm, and 54 ± 2 pm from Pr, Sc, and O, respectively. By comparing the measured column widths with the Gaussian convolved potential, we can obtain the combined broadening factors (i.e., the FWHMs of the convolved Gaussian) of 28, 25, and 34 pm for Pr, Sc, and O, respectively (fig. S12 and table S1).
Fig. 3 Spatial resolution and measurement precision from multislice electron ptychography.
(A) Projected electrostatic potential of PrScO3 in the absence of thermal vibrations. (B) Projected potential after the thermal broadening at 300 K. (C) Cropped region of the ptychographically reconstructed potential from simulated data, including thermal fluctuations from a 21-nm-thick PrScO3 sample. (D) Cropped region of the ptychographically reconstructed potential from experimental data. The potential in both (C) and (D) is normalized to the unit-cell thickness (8.03 Å). (E) Histogram of projected Pr–Pr distances from experimental measurements. The average Pr–Pr distance is 0.586 ± 0.007 Å. (F) Histogram of the projected Sc–O bond length from experimental measurements. d1 and d2 indicate different Sc–O distances along the two vertical directions, as labeled in (D). The average Sc–O distances for d1 and d2 are 2.030 ± 0.015 Å and 2.052 ± 0.013 Å, respectively. Scale bars in (A) to (D), 1 Å. st.d., standard deviation.
Thermal broadening factors of atoms can be calculated from Debye-Waller factors (DWFs) obtained from x-ray diffraction (XRD) measurements. We also measured DWFs from quantitative convergent beam electron diffraction (QCBED) (table S2 and fig. S17) (29). The DWFs of Pr and Sc atoms from our QCBED measurements agree well with those from XRD measurements of PrScO3 single crystals (26). For example, after converting to the FWHM of the thermal displacement, the thermal broadening factor for Pr from both XRD (26) and QCBED is 23 pm, compared to the measured total broadening of 28 pm for our reconstructed image. In other words, most of the measured broadening is already accounted for by the thermal vibrations in the sample. The remaining residual broadening can be attributed primarily to the uncertainty imposed by the finite incident dose and uncorrected instabilities (15). We estimate the residual instrumental contribution via quadratic subtraction, which gives the residual blurring (as a Gaussian FWHM) of our ptychographic reconstruction at Pr and Sc sites of 16 ± 1 pm and 15 ± 1 pm, respectively. Notably, QCBED produces a larger DWF for oxygen site #2 (O#2) than does XRD. The resolution of O#2 is 23 ± 2 pm or 19 ± 2 pm, depending on whether the QCBED or XRD result is adopted. It is not surprising that the resolution estimated from different elements is different, because the quality of the ptychographic reconstruction at a finite illuminated dose is dependent on the scattering power of the object (15). Therefore, the Abbe resolution of the ptychographic reconstruction, which approximately corresponds to the FWHM of the point-spread function (23), is better than 15 ± 1 pm, and its Rayleigh criteria, with a factor of 1.22 correspondence, is 18 ± 1 pm. In all cases, the column width of the ptychographic reconstruction is mainly limited by the finite size of atoms, as determined by their thermal fluctuations instead of the imaging system itself. Going forward, this method should be capable of measuring directional anisotropies of thermal vibrations for individual point-defect–like structures such as nitrogen-vacancy centers in diamond, which are being explored for applications in quantum computing and quantum communications.
In addition to the resolution improvement, the precision for measuring the atomic positions is also significantly improved. Figure 3E shows repeated measurements of Pr–Pr atomic distances with a standard deviation of 0.7 pm, which indicates that we have achieved a subpicometer precision simultaneously with the 16-pm resolution. Notably, the positions of the light oxygen are also measured precisely. Because there are two different bond lengths in the distorted ScO6 octahedra in PrScO3, the Sc–O distances along two vertical directions, labeled as d1 and d2 in Fig. 3D, are not equal. The precision is close to 1 pm, and the histogram in Fig. 3F shows d1 and d2 values of 203.0 ± 1.5 pm and 205.2 ± 1.3 pm, respectively. d1 and d2 differ by only ~2 pm, and picometer precision is required to distinguish such a small difference. However, high precision of atomic positions from conventional imaging techniques can be realized for heavy metals only (30); it cannot be achieved reliably for light elements such as oxygen (31). High precision of atomic positions and high-resolution measurement of both heavy and light atoms are crucial for correlating structure and function in materials. Multislice ptychography does not rely on lateral periodicity, so this technique can be applied to defects and grain boundaries as well as to single crystals.
Multislice electron ptychography also allows for 3D structure determination, because it iteratively retrieves the sample structure at different layers. The depth resolution and optimal depth sampling are determined by the out-of-plane curvature of the Ewald sphere and the largest scattering vector at which usable information can be collected (21). First, we check the reconstructed structures at different slices from the experimental results; three example slices are shown in Fig. 4A (all slices depicted in movie S1). We find that the slices at the beginning and the end (movie S1) show very small phase shifts. In the middle slices, the phase images show strong contrast and clear structural features. Figure 4B illustrates a depth profile of the phase change cut along the Pr–O direction (marked by the dash-dotted line in Fig. 4A). This depth evolution is generated because the sample is close to a parallel-sided film and there are vacuum layers above and beneath the film (Fig. 4C). The electron beam changes shape as a function of depth into the sample, which is properly accounted for in the multislice electron ptychography algorithm, and as a result recovers a phase shift that is linearly proportional to the electrostatic potential of the sample at each different depth slice. The broadening of the depth profile from each surface of the rectangular slab is fitted by an error function (23) and gives a depth resolution of ~3.9 nm, estimated from the width of the error function fitted to the phase variation versus depth at Pr sites (Fig. 4D). This value is better than the aperture-limited depth resolution of 5.1 nm from conventional optical sectioning (23).
Fig. 4 Depth sectioning of multislice electron ptychography.
(A) Ptychographically reconstructed phase of three selected slices at different depths (z) from experimental measurements of PrScO3. The sample thickness is initially assumed as 30 nm during the reconstruction, but the retrieved sample thickness is 21 nm. (B) Depth profile of phase intensity along the Pr–O direction marked in (A) by the blue dash-dotted line. (C) Cartoon representation of the electron beam spreading into the sample. (D) Measured depth profile of the phase at the different atomic columns indicated in (A). FW80Ms from the error function fittings are 3.9, 3.5, and 4.2 nm of Pr, Sc, and O sites, respectively. (E) Simulated ptychographic reconstructions of the phase for three selected slices from a model with two interstitial Pr atom dopants located at a depth separation of 3 nm. The dash-dotted line indicates the lateral position for the depth profile in (F); arrows denote the position of the Pr dopants. (F) Depth variation of phase intensity across the Pr dopants. The two Pr dopants are marked with arrows. (G) Depth variation of the phase for different dopant positions. Solid lines are Gaussian fits. The results in (E) to (G) are for a modeled illuminated dose of 108 e·Å−2 (e, electrons). (H) Computed depth resolution for multislice electron ptychography as a function of illuminated dose. The aperture-limited depth resolution (4.3 nm) is shown as a dashed horizontal line. Scale bars in (A) and (E), 2 Å.
We also find, through simulations, that multislice electron ptychography can enable the detection and location of interstitials and single atomic dopants in all three dimensions. We constructed a structural model by introducing single dopants in a 15-nm-thick PrScO3 crystalline matrix and generated diffraction data by using mutislice simulations (23). Dopant atoms with different atomic numbers (Z) will exhibit different visibilities in multislice electron ptychography because their potentials vary in width and magnitude. Therefore, we introduced defect atoms with a range of scattering strengths, from heavy (Pr; Z = 59) to light (O; Z = 8) and placed them at both interstitial (Fig. 4, E and F) and substitutional (fig. S14) sites and at different depth positions. Figure 4E shows the reconstructed phase images from 3 of 30 slices (movie S2) from the two Pr dopants at depths that differ by 3 nm. The depth profile across the two single dopants (Fig. 4F) shows well-localized contrast in all three dimensions. The depth resolution estimated from the full width at 80% of the maximum (FW80M) at the dopant peak in the depth profile (Pr2 in Fig. 4G) is 0.9 nm. Depth profiles from single Pr, Sc, and O dopants (Fig. 4G) are similar and show FW80M in depths of 0.9, 1.6, and 2.0 nm, respectively. This also shows that the depth resolution depends on the scattering power of the dopant, which in turn depends on the atomic number, as noted for the lateral resolution discussed earlier. A strong contrast and good depth resolution are also retained for single substitutional dopants on the atomic columns (fig. S14). Notably, the depth resolution from multislice electron ptychography exceeds the aperture-limited resolution ( λ α 2 , where λ is the wavelength and α is the probe-forming semi-angle) and has an illumination-dose dependence (Fig. 4H). Additionally, the depth resolution depends on contrast and transverse resolution and can be potentially improved by collecting electrons scattered to higher angles and using a more converging or diverging electron beam (21).
Because conventional imaging methods are based on projections through the entire sample, single dopants embedded in a relatively thick matrix usually show little or no contrast (fig. S16). Only dopants with large atomic numbers in a light element matrix and a very thin sample may be detectable. For example, an Sb atom in a silicon matrix can be distinguished by STEM annular dark-field imaging only in samples thinner than 5 nm (32). Furthermore, there is strong multiple electron scattering in crystalline samples, and the resultant rechanneling of the electron beam may lead to incorrect column locations for dopants via conventional optical sectioning methods (33). These artifacts also have consequences for electron tomography, the most commonly used technique for determining 3D structures of samples, because poor or incorrect contrast from single dopants in the projection images used for electron tomography hinders their identification and correct localization. The recovered linear signal provided by our full inversion of electron multiple scattering via multislice electron ptychography addresses these deficiencies. Therefore, multislice electron ptychography provides a powerful tool for locating single dopants at more than twice the resolution in all three dimensions compared with conventional approaches. Combined with tomography, robust atomic-resolution details of defect clusters should be possible in all three dimensions.

Acknowledgments

D.A.M. thanks M. Humphry of Phasefocus for helpful discussions on multislice 3D ptychography. Funding: This research was supported by the National Science Foundation under grant DMR-1539918 (PARADIM Materials Innovation Platform in-house program). Electron microscopy performed at the Cornell Center for Materials Research was supported by the National Science Foundation under grant DMR-1719875. Author contributions: Z.C. and D.A.M. initiated and conceived the research; Z.C. performed the experiments and data analyses under the supervision of D.A.M.; Z.C. and Y.J. implemented the mixed-state multislice ptychography algorithms based on the initial ptychography codes from M.O. and M.G.-S.; Y.-T.S. performed the Debye-Waller factors measurements; M.E.H. prepared the sample under the supervision of D.G.S., who also provided advice on XRD structural refinements; I.H. grew the PrScO3 single crystal by the Czochralski method, under the supervision of S.G.; and Z.C. wrote the manuscript, with revisions from D.A.M. and input from all authors. Competing interests: Cornell University has licensed the EMPAD hardware to Thermo Fisher Scientific. Data and materials availability: Raw experimental data of the results presented in the main text and supplementary materials are available from PARADIM, a National Science Foundation Materials Innovation Platform (34). The source code for multislice electron ptychography is available in the Zenodo repository (35).

Supplementary Material

Summary

Materials and Methods
Figs. S1 to S17
Tables S1 and S2
References (3658)
Movies S1 and S2

Resources

File (abg2533_chen_sm.pdf)
File (abg2533s1.mp4)
File (abg2533s2.mp4)

References and Notes

1
M. Haider, S. Uhlemann, E. Schwan, H. Rose, B. Kabius, K. Urban, Electron microscopy image enhanced. Nature 392, 768–769 (1998).
2
P. E. Batson, N. Dellby, O. L. Krivanek, Sub-ångstrom resolution using aberration corrected electron optics. Nature 418, 617–620 (2002).
3
R. Erni, M. D. Rossell, C. Kisielowski, U. Dahmen, Atomic-resolution imaging with a sub-50-pm electron probe. Phys. Rev. Lett. 102, 096101 (2009).
4
H. Sawada, Y. Tanishiro, N. Ohashi, T. Tomita, F. Hosokawa, T. Kaneyama, Y. Kondo, K. Takayanagi, STEM imaging of 47-pm-separated atomic columns by a spherical aberration-corrected electron microscope with a 300-kV cold field emission gun. J. Electron Microsc. 58, 357–361 (2009).
5
J. M. Cowley, A. F. Moodie, The scattering of electrons by atoms and crystals. I. A new theoretical approach. Acta Cryst. 10, 609–619 (1957).
6
M. J. Hÿtch, W. M. Stobbs, Quantitative comparison of high resolution TEM images with image simulations. Ultramicroscopy 53, 191–203 (1994).
7
H. G. Brown, Z. Chen, M. Weyland, C. Ophus, J. Ciston, L. J. Allen, S. D. Findlay, Structure Retrieval at Atomic Resolution in the Presence of Multiple Scattering of the Electron Probe. Phys. Rev. Lett. 121, 266102 (2018).
8
J. J. Donatelli, J. C. H. Spence, Inversion of Many-Beam Bragg Intensities for Phasing by Iterated Projections: Removal of Multiple Scattering Artifacts from Diffraction Data. Phys. Rev. Lett. 125, 065502 (2020).
9
W. Hoppe, Beugung im inhomogenen Primarstrahlwellenfeld. I. Prinzip einer Phasenmessung von Elektronenbeungungsinterferenzen. Acta Cryst. A 25, 495–501 (1969).
10
J. M. Rodenburg, H. M. L. Faulkner, A phase retrieval algorithm for shifting illumination. Appl. Phys. Lett. 85, 4795–4797 (2004).
11
J. M. Rodenburg, A. C. Hurst, A. G. Cullis, Transmission microscopy without lenses for objects of unlimited size. Ultramicroscopy 107, 227–231 (2007).
12
P. Thibault, M. Dierolf, A. Menzel, O. Bunk, C. David, F. Pfeiffer, High-resolution scanning x-ray diffraction microscopy. Science 321, 379–382 (2008).
13
T. J. Pennycook, A. R. Lupini, H. Yang, M. F. Murfitt, L. Jones, P. D. Nellist, Efficient phase contrast imaging in STEM using a pixelated detector. Part 1: Experimental demonstration at atomic resolution. Ultramicroscopy 151, 160–167 (2015).
14
S. Gao, P. Wang, F. Zhang, G. T. Martinez, P. D. Nellist, X. Pan, A. I. Kirkland, Electron ptychographic microscopy for three-dimensional imaging. Nat. Commun. 8, 163 (2017).
15
Y. Jiang, Z. Chen, Y. Han, P. Deb, H. Gao, S. Xie, P. Purohit, M. W. Tate, J. Park, S. M. Gruner, V. Elser, D. A. Muller, Electron ptychography of 2D materials to deep sub-ångström resolution. Nature 559, 343–349 (2018).
16
R. Close, Z. Chen, N. Shibata, S. D. Findlay, Towards quantitative, atomic-resolution reconstruction of the electrostatic potential via differential phase contrast using electrons. Ultramicroscopy 159, 124–137 (2015).
17
Z. Chen, M. Odstrcil, Y. Jiang, Y. Han, M.-H. Chiu, L.-J. Li, D. A. Muller, Mixed-state electron ptychography enables sub-angstrom resolution imaging with picometer precision at low dose. Nat. Commun. 11, 2994 (2020).
18
A. M. Maiden, M. J. Humphry, J. M. Rodenburg, Ptychographic transmission microscopy in three dimensions using a multi-slice approach. J. Opt. Soc. Am. A 29, 1606–1614 (2012).
19
T. M. Godden, R. Suman, M. J. Humphry, J. M. Rodenburg, A. M. Maiden, Ptychographic microscope for three-dimensional imaging. Opt. Express 22, 12513–12523 (2014).
20
A. Suzuki, S. Furutaku, K. Shimomura, K. Yamauchi, Y. Kohmura, T. Ishikawa, Y. Takahashi, High-resolution multislice x-ray ptychography of extended thick objects. Phys. Rev. Lett. 112, 053903 (2014).
21
E. H. Tsai, I. Usov, A. Diaz, A. Menzel, M. Guizar-Sicairos, X-ray ptychography with extended depth of field. Opt. Express 24, 29089–29108 (2016).
22
M. Schloz, T. C. Pekin, Z. Chen, W. Van den Broek, D. A. Muller, C. T. Koch, Overcoming information reduced data and experimentally uncertain parameters in ptychography with regularized optimization. Opt. Express 28, 28306–28323 (2020).
23
Materials and methods are available as supplementary materials.
24
P. A. Midgley, M. Weyland, 3D electron microscopy in the physical sciences: The development of Z-contrast and EFTEM tomography. Ultramicroscopy 96, 413–431 (2003).
25
M. W. Tate, P. Purohit, D. Chamberlain, K. X. Nguyen, R. Hovden, C. S. Chang, P. Deb, E. Turgut, J. T. Heron, D. G. Schlom, D. C. Ralph, G. D. Fuchs, K. S. Shanks, H. T. Philipp, D. A. Muller, S. M. Gruner, High dynamic range pixel array detector for scanning transmission electron microscopy. Microsc. Microanal. 22, 237–249 (2016).
26
T. M. Gesing, R. Uecker, J. C. Buhl, Refinement of the crystal structure of praseodymium orthoscandate, PrScO3. Z. Kristallogr. New Cryst. Struct. 224, 365–366 (2009).
27
P. Thibault, A. Menzel, Reconstructing state mixtures from diffraction measurements. Nature 494, 68–71 (2013).
28
M. A. O’Keefe, Seeing atoms with aberration-corrected sub-Angström electron microscopy. Ultramicroscopy 108, 196–209 (2008).
29
J. M. Zuo, Automated lattice parameter measurement from HOLZ lines and their use for the measurement of oxygen content in YBa2Cu3O7-δ from nanometer-sized region. Ultramicroscopy 41, 211–223 (1992).
30
A. B. Yankovich, B. Berkels, W. Dahmen, P. Binev, S. I. Sanchez, S. A. Bradley, A. Li, I. Szlufarska, P. M. Voyles, Picometre-precision analysis of scanning transmission electron microscopy images of platinum nanocatalysts. Nat. Commun. 5, 4155 (2014).
31
D. Zhou, K. Müller-Caspary, W. Sigle, F. F. Krause, A. Rosenauer, P. A. van Aken, Sample tilt effects on atom column position determination in ABF-STEM imaging. Ultramicroscopy 160, 110–117 (2016).
32
P. M. Voyles, D. A. Muller, J. L. Grazul, P. H. Citrin, H. J. L. Gossmann, Atomic-scale imaging of individual dopant atoms and clusters in highly n-type bulk Si. Nature 416, 826–829 (2002).
33
H. L. Xin, V. Intaraprasonk, D. A. Muller, Depth sectioning of individual dopant atoms with aberration-corrected scanning transmission electron microscopy. Appl. Phys. Lett. 92, 013125 (2008).
34
Z. Chen, Y. Jiang, Y.-T. Shao, M. E. Holtz, M. Odstrčil, M. Guizar-Sicairos, I. Hanke, S. Ganschow, D. G. Schlom, D. A. Muller, Data Set: Electron ptychography achieves atomic-resolution limits set by lattice vibrations, PARADIM (2021), https://doi.org/10.34863/ssmm-2j11.
35
Z. Chen, Y. Jiang, D. A. Muller, M. Odstrcil, PtychoShelves_EM, source code for multislice electron ptychography, version 2, Zenodo (2021); https://doi.org/10.5281/zenodo.4659690.
36
R. Uecker, H. Wilke, D. G. Schlom, B. Velickov, P. Reiche, A. Polity, M. Bernhagen, M. Rossberg, Spiral formation during Czochralski growth of rare-earth scandates. J. Cryst. Growth 295, 84–91 (2006).
37
J. M. Rodenburg, R. H. T. Bates, The theory of super-resolution electron microscopy via Wigner-distribution deconvolution. Philos. Trans. R. Soc. London Ser. A 339, 521–553 (1992).
38
M. Odstrčil, A. Menzel, M. Guizar-Sicairos, Iterative least-squares solver for generalized maximum-likelihood ptychography. Opt. Express 26, 3108–3123 (2018).
39
K. Wakonig, H.-C. Stadler, M. Odstrcil, E. H. R. Tsai, A. Diaz, M. Holler, I. Usov, J. Raabe, A. Menzel, M. Guizar-Sicairos, PtychoShelves, a versatile high-level framework for high-performance analysis of ptychographic data. J. Appl. Cryst. 53, 574–586 (2020).
40
P. M. Pelz, W. X. Qiu, R. Bücker, G. Kassier, R. J. D. Miller, Low-dose cryo electron ptychography via non-convex Bayesian optimization. Sci. Rep. 7, 9883 (2017).
41
J. Song, C. S. Allen, S. Gao, C. Huang, H. Sawada, X. Pan, J. Warner, P. Wang, A. I. Kirkland, Atomic resolution defocused electron ptychography at low dose with a fast, direct electron detector. Sci. Rep. 9, 3919 (2019).
42
L. Zhou, J. Song, J. S. Kim, X. Pei, C. Huang, M. Boyce, L. Mendonça, D. Clare, A. Siebert, C. S. Allen, E. Liberti, D. Stuart, X. Pan, P. D. Nellist, P. Zhang, A. I. Kirkland, P. Wang, Low-dose phase retrieval of biological specimens using cryo-electron ptychography. Nat. Commun. 11, 2773 (2020).
43
H. Yang, I. MacLaren, L. Jones, G. T. Martinez, M. Simson, M. Huth, H. Ryll, H. Soltau, R. Sagawa, Y. Kondo, C. Ophus, P. Ercius, L. Jin, A. Kovács, P. D. Nellist, Electron ptychographic phase imaging of light elements in crystalline materials using Wigner distribution deconvolution. Ultramicroscopy 180, 173–179 (2017).
44
Y. Jiang, Z. Chen, I. El Baggari, L. F. Kourkoutis, V. Elser, D. A. Muller, Breaking the Rayleigh limit in thick samples with multi-slice ptychography. Microsc. Microanal. 24 (suppl. 1), 192–193 (2018).
45
C. Ophus, P. Ercius, M. Sarahan, C. Czarnik, J. Ciston, Recording and Using 4D-STEM Datasets in Materials Science. Microsc. Microanal. 20 (suppl. 3), 62–63 (2014).
46
J. A. Mir, R. Clough, R. MacInnes, C. Gough, R. Plackett, I. Shipsey, H. Sawada, I. MacLaren, R. Ballabriga, D. Maneuski, V. O’Shea, D. McGrouther, A. I. Kirkland, Characterisation of the Medipix3 detector for 60 and 80keV electrons. Ultramicroscopy 182, 44–53 (2017).
47
E. J. Kirkland, Advanced Computing in Electron Microscopy (Springer, ed. 2, 2010).
48
P. A. Doyle, P. S. Turner, Relativistic Hartree-Fock X-ray and electron scattering factors. Acta Cryst. A 24, 390–397 (1968).
49
A. Weickenmeier, H. Kohl, Computation of Absorptive form factors for high-energy electron diffraction. Acta Cryst. A 47, 590–597 (1991).
50
B. Jiang, J.-M. Zuo, J. Friis, J. C. H. Spence, On the consistency of QCBED structure factor measurements for TiO2 (rutile). Microsc. Microanal. 9, 457–467 (2003).
51
J. M. Zuo, J. C. H. Spence, Electron Microdiffraction (Springer, 1992).
52
J. M. Zuo, Quantitative Convergent Beam Electron Diffraction. Mater. Trans. JIM 39, 938–946 (1998).
53
R. P. Liferovich, R. H. Mitchell, A structural study of ternary lanthanide orthoscandate perovskites. J. Solid State Chem. 177, 2188–2197 (2004).
54
L. Reimer, Transmission Electron Microscopy (Springer, 1989).
55
L. J. Allen, A. J. D’Alfonso, S. D. Findlay, Modelling the inelastic scattering of fast electrons. Ultramicroscopy 151, 11–22 (2015).
56
M. Born, E. Wolf, Principles of Optics: Electromagnetic Theory of Propagation, Interference, and Diffraction of Light (Pergamon, 1989).
57
H. L. Xin, D. A. Muller, Aberration-corrected ADF-STEM depth sectioning and prospects for reliable 3D imaging in S/TEM. J. Electron Microsc. 58, 157–165 (2009).
58
C.-L. Jia, M. Lentzen, K. Urban, High-resolution transmission electron microscopy using negative spherical aberration. Microsc. Microanal. 10, 174–184 (2004).

(0)eLetters

eLetters is a forum for ongoing peer review. eLetters are not edited, proofread, or indexed, but they are screened. eLetters should provide substantive and scholarly commentary on the article. Embedded figures cannot be submitted, and we discourage the use of figures within eLetters in general. If a figure is essential, please include a link to the figure within the text of the eLetter. Please read our Terms of Service before submitting an eLetter.

Log In to Submit a Response

No eLetters have been published for this article yet.

Information & Authors

Information

Published In

Science
Volume 372 | Issue 6544
21 May 2021

Submission history

Received: 21 December 2020
Accepted: 13 April 2021
Published in print: 21 May 2021

Permissions

Request permissions for this article.

Acknowledgments

D.A.M. thanks M. Humphry of Phasefocus for helpful discussions on multislice 3D ptychography. Funding: This research was supported by the National Science Foundation under grant DMR-1539918 (PARADIM Materials Innovation Platform in-house program). Electron microscopy performed at the Cornell Center for Materials Research was supported by the National Science Foundation under grant DMR-1719875. Author contributions: Z.C. and D.A.M. initiated and conceived the research; Z.C. performed the experiments and data analyses under the supervision of D.A.M.; Z.C. and Y.J. implemented the mixed-state multislice ptychography algorithms based on the initial ptychography codes from M.O. and M.G.-S.; Y.-T.S. performed the Debye-Waller factors measurements; M.E.H. prepared the sample under the supervision of D.G.S., who also provided advice on XRD structural refinements; I.H. grew the PrScO3 single crystal by the Czochralski method, under the supervision of S.G.; and Z.C. wrote the manuscript, with revisions from D.A.M. and input from all authors. Competing interests: Cornell University has licensed the EMPAD hardware to Thermo Fisher Scientific. Data and materials availability: Raw experimental data of the results presented in the main text and supplementary materials are available from PARADIM, a National Science Foundation Materials Innovation Platform (34). The source code for multislice electron ptychography is available in the Zenodo repository (35).

Authors

Affiliations

School of Applied and Engineering Physics, Cornell University, Ithaca, NY 14853, USA.
Advanced Photon Source, Argonne National Laboratory, Lemont, IL 60439, USA.
Yu-Tsun Shao
School of Applied and Engineering Physics, Cornell University, Ithaca, NY 14853, USA.
Department of Materials Science and Engineering, Cornell University, Ithaca, NY, USA.
Present address: Department of Metallurgical and Materials Engineering, Colorado School of Mines, Golden, CO 80401, USA.
Paul Scherrer Institut, 5232 Villigen PSI, Switzerland.
Present address: Carl Zeiss SMT, Carl-Zeiss-Straße 22, 73447 Oberkochen, Germany.
Manuel Guizar-Sicairos https://orcid.org/0000-0002-8293-3634
Paul Scherrer Institut, 5232 Villigen PSI, Switzerland.
Isabelle Hanke
Leibniz-Institut für Kristallzüchtung, Max-Born-Str. 2, 12489 Berlin, Germany.
Leibniz-Institut für Kristallzüchtung, Max-Born-Str. 2, 12489 Berlin, Germany.
Department of Materials Science and Engineering, Cornell University, Ithaca, NY, USA.
Leibniz-Institut für Kristallzüchtung, Max-Born-Str. 2, 12489 Berlin, Germany.
Kavli Institute at Cornell for Nanoscale Science, Ithaca, NY, USA.
School of Applied and Engineering Physics, Cornell University, Ithaca, NY 14853, USA.
Kavli Institute at Cornell for Nanoscale Science, Ithaca, NY, USA.

Funding Information

Notes

*
Corresponding author. Email: [email protected] (Z.C.); [email protected] (D.A.M.)

Metrics & Citations

Metrics

Article Usage

Altmetrics

Citations

Cite as

Export citation

Select the format you want to export the citation of this publication.

Cited by

  1. Achieving sub-0.5-angstrom–resolution ptychography in an uncorrected electron microscope, Science, 383, 6685, (865-870), (2024)./doi/10.1126/science.adl2029
    Abstract
  2. Chemical Electron Microscopy (CEM) for Heterogeneous Catalysis at Nano: Recent Progress and Challenges, Research, 6, (2023)./doi/10.34133/research.0043
    Abstract
  3. Probing the Boundary between Classical and Quantum Mechanics by Analyzing the Energy Dependence of Single-Electron Scattering Events at the Nanoscale, Nanomaterials, 13, 6, (971), (2023).https://doi.org/10.3390/nano13060971
    Crossref
  4. Three-dimensional inhomogeneity of zeolite structure and composition revealed by electron ptychography, Science, 380, 6645, (633-638), (2023)./doi/10.1126/science.adg3183
    Abstract
  5. Ptychographic measurements of varying size and shape along zeolite channels, Science Advances, 9, 11, (2023)./doi/10.1126/sciadv.adf1151
    Abstract
  6. Deep-Learning Electron Diffractive Imaging, Physical Review Letters, 130, 1, (2023).https://doi.org/10.1103/PhysRevLett.130.016101
    Crossref
  7. Direct observation of Cu in high-silica chabazite zeolite by electron ptychography using Wigner distribution deconvolution, Scientific Reports, 13, 1, (2023).https://doi.org/10.1038/s41598-023-27452-3
    Crossref
  8. Operando studies reveal active Cu nanograins for CO2 electroreduction, Nature, 614, 7947, (262-269), (2023).https://doi.org/10.1038/s41586-022-05540-0
    Crossref
  9. Time-resolved transmission electron microscopy for nanoscale chemical dynamics, Nature Reviews Chemistry, 7, 4, (256-272), (2023).https://doi.org/10.1038/s41570-023-00469-y
    Crossref
  10. Sub-nanometer-scale mapping of crystal orientation and depth-dependent structure of dislocation cores in SrTiO3, Nature Communications, 14, 1, (2023).https://doi.org/10.1038/s41467-023-35877-7
    Crossref
  11. See more
Loading...

View Options

View options

PDF format

Download this article as a PDF file

Download PDF

Check Access

Log in to view the full text

AAAS ID LOGIN

AAAS login provides access to Science for AAAS Members, and access to other journals in the Science family to users who have purchased individual subscriptions.

Log in via OpenAthens.
Log in via Shibboleth.

More options

Register for free to read this article

As a service to the community, this article is available for free. Login or register for free to read this article.

Purchase this issue in print

Buy a single issue of Science for just $15 USD.

Media

Figures

Multimedia

Tables

Share

Share

Share article link

Share on social media