Brief Report
Open access
Published Online: 1 October 2017

FlexTouch: An Insulin Pen-Injector with a Low Activation Force Across Different Insulin Formulations, Needle Technologies, and Temperature Conditions

Publication: Diabetes Technology & Therapeutics
Volume 19, Issue Number 10

Abstract

FlexTouch® (Novo Nordisk A/S, Bagsvaerd, Denmark) is a pen-injector with a torque spring mechanism requiring a low activation force. This laboratory-based study compared the activation force of FlexTouch during the injection of insulin with different needles and at temperature conditions within the range at which the device is recommended for use. Using a tensile tester, activation force was measured at maximum dose settings for insulin detemir (100 U/mL) and insulin degludec (100 and 200 U/mL) at standard (23°C ± 5°C), cool (5°C ± 3°C), and warm (30°C ± 2°C) conditions. Activation force was measured with two 32-gauge needles differing in internal diameter at standard conditions. At standard, cool, and warm conditions, estimated mean activation forces with 95% confidence interval were 5.71 newtons (N) (5.63–5.79), 5.94 N (5.83–6.06), and 5.69 N (5.58–5.80) with insulin detemir, 5.53 N (5.45–5.62), 5.56 N (5.44–5.67), and 5.33 N (5.22–5.44) with 100 U/mL insulin degludec, and 5.53 N (5.45–5.61), 5.83 N (5.71–5.94), and 5.56 N (5.45–5.68) with 200 U/mL insulin degludec, respectively. Mean activation forces were observed to be low with very small variability between measurements; however, the differences between insulins and temperature conditions were statistically significant. The activation force required by FlexTouch remained low across all situations tested. The differences between activation force needed with different insulins and temperature conditions were small and unlikely to be clinically meaningful.

Introduction

Insulin is essential in the treatment of type 1 diabetes (T1D) and type 2 diabetes (T2D) if other therapies become inadequate. Insulin pen-injectors are now widely used for insulin delivery as they offer significant benefits over traditional vials and syringes, including better dose accuracy and consistency, improved treatment regimen adherence and persistence, and lower HbA1c levels.1–4 Pen-injectors are also preferred by the majority of patients for their greater ease of use, confidence in dosing accuracy, and lower injection pain.1,5–8
Notwithstanding these benefits, current limitations associated with pen-injectors include the extension of the dose button (especially as the dose increases), handling during injection, and the manual force needed to deliver the insulin.
Most pen-injectors rely on thumb pressure to depress an extended dose button (termed dose force). This force needs to be applied throughout the entire injection and directly influences the rate of insulin delivery. Depressing the dose button can be particularly difficult for younger patients who cannot reach over the extension of the dose button and patients who may have impaired manual dexterity, strength, and nerve function, for example, elderly patients.9,10
FlexTouch® (Novo Nordisk A/S, Denmark) is a pen-injector with a torque spring injection mechanism and no dose button extension.11 The activation force (equivalent to the dose force for pen-injectors with extending dose button) is loaded into the internal spring when the dose is set, and the spring's energy is activated when the dose button is pressed. As a consequence, the injection force of FlexTouch is significantly lower than other widely used prefilled pens.12,13
It is important that FlexTouch retains a low and consistent activation force during everyday use. Variables that may affect the functioning of FlexTouch include the inner diameter of the needle being used and the external temperature.14,15 The aim of this study was to determine the activation force of FlexTouch with maximum doses of commonly used insulins, two different needles, and at temperature conditions within the range at which the device is recommended for use.

Materials and Methods

Materials

The insulin formulations tested were insulin detemir (100 U/mL; batch EP50929), insulin degludec (100 U/mL; batch EP50570), and insulin degludec (200 U/mL; batch EP50955). Two 32-gauge needles with different internal diameters were used (NovoFine Plus®, Novo Nordisk, 4 mm, extra thin wall [ETW], batch 14F04N; SureComfort®, Allison Medical, Inc., USA, 6 mm, thin wall [TW], batch P140712).

Equipment

A climate chamber (Vötsch WKL34; Vötsch Industrietechnik GmbH, Balingen, Germany) was used to establish three temperature conditions: standard (23°C ± 5°C with a relative humidity [RH] of 50% ± 25%), cool (5°C ± 3°C with no RH requirement), and warm (30°C ± 2°C with a 50% ± 10% RH). Activation force was measured in newtons (N) using a universal tensile tester (Lloyd LRX; Lloyd Instruments Ltd., Humlebaek, Denmark), with a 100 N load cell.

Activation force testing

Laboratory testing was performed at the SP Technical Research Institute of Sweden (Sp Sveriges Tekniska Forskningsinstitut A/S, Denmark). Activation force was measured at the maximum injectable dose setting: 80 U for insulin detemir (100 U/mL) and insulin degludec (100 U/mL), and 160 U for insulin degludec (200 U/mL) with ETW and TW needles. The insulin formulations were tested with the ETW needle at standard, cool, and warm conditions and with the TW needle at standard conditions.
Needles and pen-injectors were acclimatized for at least 4 h in the climate chamber. One pen-injector at a time was taken out of the chamber and used for testing within 3 min. A needle was attached to the pen-injector, and air expelled until liquid appeared at the tip of the needle. The dose was set and the pen-injector was then placed in the tensile tester with the needle facing downward. The tensile tester was set to a preload of 0.2 N and an approaching speed of 0.5 mm/s. After preload was reached, force measurements were performed at a pressing speed of 1 mm/s until a distance of 2.8 mm from the position where preload was reached. The activation of the injection mechanism was indicated by a peak in the force readings after pressing ∼1.4 mm and this was recorded as the activation force. For each experimental condition, a mean activation force was obtained from 25 measurements using 25 pen-injectors, with 1 measurement per pen-injector. A new needle was used for each measurement.

Statistics

Differences in activation force between insulin detemir (100 U/mL) and insulin degludec (100 and 200 U/mL), temperature condition, and needle were investigated by a linear normal model (analysis of variance) with the needle and the interaction between insulin and temperature condition as fixed effects. The estimated means and estimated differences between means with 95% confidence intervals (CI) were calculated.

Results

The estimated mean activation forces measured with insulin detemir (100 U/mL) and insulin degludec (100 and 200 U/mL) under different temperature conditions are shown in Figure 1. Although the differences in the mean activation forces for each combination of insulin and temperature condition were statistically significant (P < 0.0001), these differences were small and are unlikely to be of clinical relevance. Mean activation forces were between 5 and 6 N with an ∼5% decrease in the estimated mean activation force across the cool, standard, and warm temperature conditions (estimated differences between temperatures [95% CI]; insulin detemir: cool vs. standard 0.23 N [0.09–0.37], standard vs. warm 0.02 N [−0.12 to 0.16]; insulin degludec [100 U/mL]: cool vs. standard 0.02 N [−0.12 to 0.16], standard vs. warm 0.20 N [0.06–0.34]; insulin degludec [200 U/mL]: cool vs. standard 0.30 N [0.16–0.44], standard vs. warm −0.03 N [−0.17 to 0.11]). The mean activation forces were similar between the 32-gauge ETW needles (5.62 N [95% CI: 5.58–5.66]) and TW needles (5.70 N [95% CI: 5.61–5.78]) when accounting for differences with temperature condition and insulin (estimated difference between needles [95% CI]: −0.08 N [−0.17–0.01]; P = 0.10).
FIG. 1. Estimated activation force by compound and temperature. Activation force required by FlexTouch® to inject maximum doses of insulin detemir (100 U/mL) and insulin degludec (100 and 200 U/mL) under different temperature conditions. Data are shown as least square means (95% confidence intervals) estimated from a linear normal model. N, newtons; U, units.

Discussion

FlexTouch pen-injectors are currently available with a range of insulins and different needles. When not being used, pen-injectors should be kept in a refrigerator at a temperature between 2°C and 8°C and away from the freezing element. When in use, it is recommended that the FlexTouch pen-injector is operated at room temperature, below 30°C, and out of direct sunlight. However, diverse climatic conditions can result in pen-injectors being exposed to a range of temperatures during everyday use and it is important that performance is not adversely affected by these variables. This is the first study to demonstrate that the low activation force of FlexTouch is only minimally affected by temperature conditions representing the extremes of the recommended range for use. The small differences observed between conditions in these laboratory tests are unlikely to be of clinical relevance.
The torque spring injection mechanism of FlexTouch results in a lower activation force compared with the dose force of other pen-injectors (Table 1). Previous technical studies show the mean (standard deviation) activation force of FlexTouch to be 5.1 (0.5) N at maximum dose settings; this is between 56% and 86% lower than the force required by SoloSTAR® (Sanofi-Aventis, Paris, France) and KwikPen® (Eli Lilly and Co., Indianapolis)12,13 and is consistent with the activation forces measured under standard conditions in this study. There was also very little variation in the repeated force measurements, as shown by the narrow CIs and the small differences between the temperature conditions, highlighting that the activation force of FlexTouch remains consistently low.
Table 1. Characteristics of the FlexTouch Pen-Injector Compared with Other Insulin Injection Devices
Characteristic Study details Comparison with FT Reference
Lower activation force Difference in activation force at 4, 6, and 8 mm/s push-button speeds FT vs. KwikPen®: –9.4,–15.9, –23.1 N (P < 0.0001) 12
    FT vs. SoloSTAR®: –8.4,–14.0, –21.8 N (P < 0.0001)  
Higher maximum flow rate Flow analysis while dispersing an 80 U dose FT vs. SoloSTAR: 15.6 vs. 6.0–10.0 U/s 21–23
Rated as the overall preferred device and as easier to use, easier to learn/teach how to use, and easier to handle/perform injection with Patient-reported outcome questionnaires from patients with diabetes, caregivers, and/or HCPs FT vs. vials and syringes 3,24,25
    FT vs. KwikPen 11,26
    FT vs. InnoLet® 16
    FT vs. SoloSTAR 26,27
    FT vs. FlexPen® 28
FT, FlexTouch®; HCP, healthcare professional; N, newtons; s, second; U, unit.
In a clinical context, the activation/dose force required to deliver insulin by a pen-injector can be a barrier to insulin therapy and the low activation force of FlexTouch aims to reduce the difficulties of injecting insulin. Usability studies show that FlexTouch is preferred to vials and syringes and other widely used pen-injectors by people with T1D and T2D, caregivers, and healthcare professionals (HCPs) (Table 1). As an example, in a study in which 61% of 90 participants with T2D had impaired dexterity, FlexTouch was considered superior to InnoLet® (Novo Nordisk A/S, Denmark) with respect to injection procedure, handling, and confidence in correct dosing.16
The low activation force of FlexTouch is reflected in a high maximum flow rate (Table 1). While injection speed has often been considered to impact upon a patient's perception of injection pain, evidence from a randomized controlled trial demonstrates that subcutaneous injection speed had no effect on injection pain.17 Indeed, a higher injection speed may contribute to a shorter dwell time of the needle in the injection site.18 For most prefilled pen-injectors, needle choice has an impact on the force required to depress the dose button, with a larger diameter associated with reduced force.19 However, thinner needles tend to be preferred by patients as they are associated with less pain, bleeding, and bruising than thicker needles.20,21 While the outer diameters of the needles used in this study were consistent, the ETW technology allows for a larger inner diameter, and so an increased flow rate, compared with the TW needles. Nonetheless, the activation force was similar with the two 32-gauge needles tested when accounting for the differences with insulins and temperature conditions, demonstrating that a patients' needle choice (ETW vs. TW technology) does not impact upon the activation force of FlexTouch. This is not unexpected as, due to the torque spring mechanism, the activation force of FlexTouch is unlikely to be affected by needle diameter.
A limitation of this study was that measurements were performed in a controlled, laboratory environment and other factors, such as the tissue being injected into, may affect injection force when performed by patients or HCPs. However, accurately measuring force in a clinical setting would be very challenging. Clinical studies are needed to assess the range over which injection force influences patient perception and preference, and if injection force is an important factor in insulin therapy adherence and, ultimately, in improved glycemic control.
In conclusion, the low activation force of FlexTouch with commonly used insulins is maintained across needle technologies and temperature conditions (within the recommended range).

Acknowledgments

The authors thank the SP Technical Research Institute of Sweden for performing the measurements. The authors also acknowledge Joakim Petersen at Novo Nordisk for his technical assistance. Medical writing and submission support were provided by Helen Parker and Beverly La Ferla of Watermeadow Medical, an Ashfield Company, part of UDG Healthcare plc, funded by Novo Nordisk. Funding sources: This study was funded by Novo Nordisk A/S.

References

1.
Kadiri A, Chraibi A, Marouan F, et al.: Comparison of NovoPen 3 and syringes/vials in the acceptance of insulin therapy in NIDDM patients with secondary failure to oral hypoglycaemic agents. Diabetes Res Clin Pract 1998;41:15–23.
2.
Asakura T, Seino H, Nakano R, et al.: A comparison of the handling and accuracy of syringe and vial versus prefilled insulin pen (FlexPen). Diabetes Technol Ther 2009;11:657–661.
3.
Pfützner A, Bailey T, Campos C, et al.: Accuracy and preference assessment of prefilled insulin pen versus vial and syringe with diabetes patients, caregivers, and healthcare professionals. Curr Med Res Opin 2013;29:475–481.
4.
Xie L, Zhou S, Pinsky BW, et al.: Impact of initiating insulin glargine disposable pen versus vial/syringe on real-world glycemic outcomes and persistence among patients with type 2 diabetes mellitus in a large managed care plan: a claims database analysis. Diabetes Technol Ther 2014;16:567–575.
5.
Graff MR, McClanahan MA: Assessment by patients with diabetes mellitus of two insulin pen delivery systems versus a vial and syringe. Clin Ther 1998;20:486–496.
6.
Korytkowski M, Bell D, Jacobsen C, et al.: A multicenter, randomized, open-label, comparative, two-period crossover trial of preference, efficacy, and safety profiles of a prefilled, disposable pen and conventional vial/syringe for insulin injection in patients with type 1 or 2 diabetes mellitus. Clin Ther 2003;25:2836–2848.
7.
Oyer D, Narendran P, Qvist M, et al.: Ease of use and preference of a new versus widely available prefilled insulin pen assessed by people with diabetes, physicians and nurses. Expert Opin Drug Deliv 2011;8:1259–1269.
8.
Ramadan WH, Khreis NA, Kabbara WK: Simplicity, safety, and acceptability of insulin pen use versus the conventional vial/syringe device in patients with type 1 and type 2 diabetes mellitus in Lebanon. Patient Prefer Adherence 2015;9:517–528.
9.
Pfützner J, Hellhammer J, Musholt P, et al.: Evaluation of dexterity in insulin-treated patients with type 1 and type 2 diabetes mellitus. J Diabetes Sci Technol 2011;5:158–165.
10.
Pfützner A, Musholt PB, Schipper C, et al.: Self-assessment and objective determination of dexterity in patients with type 1 or type 2 diabetes mellitus. Curr Med Res Opin 2012;28:15–21.
11.
Bailey T, Thurman J, Niemeyer M, et al.: Usability and preference evaluation of a prefilled insulin pen with a novel injection mechanism by people with diabetes and healthcare professionals. Curr Med Res Opin 2011;27:2043–2052.
12.
Hemmingsen H, Niemeyer M, Hansen MR, et al.: A prefilled insulin pen with a novel injection mechanism and a lower injection force than other prefilled insulin pens. Diabetes Technol Ther 2011;13:1207–1211.
13.
Götzche D, Rasmussen BØ, Pedersen MT, et al.: Injection force and dose accuracy of FlexTouch for the delivery of a new basal insulin. Expert Opin Drug Deliv 2013;10:1613–1619.
14.
Aronson R, Gibney MA, Oza K, et al.: Insulin pen needles: effects of extra-thin wall needle technology on preference, confidence, and other patient ratings. Clin Ther 2013;35:923–933.
15.
Merritt D, Schreiner B, Harris S, et al.: Dose accuracy and injection force of disposable pens delivering pramlintide for the treatment of diabetes. J Diabetes Sci Technol 2010;4:1438–1446.
16.
Schipper C, Musholt P, Niemeyer M, et al.: Patient device assessment evaluation of two insulin injection devices in a mixed cohort of insulin-treated patients with type 1 or type 2 diabetes mellitus. Curr Med Res Opin 2012;28:1297–1303.
17.
Heise T, Nosek L, Dellweg S, et al.: Impact of injection speed and volume on perceived pain during subcutaneous injections into the abdomen and thigh: a single-centre, randomized controlled trial. Diabetes Obes Metab 2014;16:971–976.
18.
Yeager HK, Ignaut DA: Analysis and perspective of dosing accuracy and insulin flow rate characteristics of a new disposable insulin pen, FlexTouch, compared with SoloSTAR. J Diabetes Sci Technol 2013;7:1027–1029.
19.
Molin A, Larsen C, Lawton SA: Reduced flow resistance in insulin pen needles designed with thin wall technology. Diabetes 2002;51(Suppl 1):A475 (abstract).
20.
McKay M, Compion G, Lytzen L: A comparison of insulin injection needles on patients' perceptions of pain, handling and acceptability. Diabetes Technol Ther 2009;11:195–201.
21.
Iwanaga M, Kamoi K: Patient perceptions of injection pain and anxiety: a comparison of NovoFine 32-gauge tip 6 mm and Micro Fine Plus 31-gauge 5 mm needles. Diabetes Technol Ther 2009;11:81–86.
22.
Bohnet J, Schmitz M, Kamlot S, et al.: Dosing accuracy and insulin flow rate characteristics of a new disposable insulin pen, FlexTouch, compared with SoloSTAR. J Diabetes Sci Technol 2013;7:1021–1026.
23.
Van der Burg T: Injection force of SoloSTAR® compared with other disposable insulin pen devices at constant volume flow rates. J Diabetes Sci Technol 2011;5:150–155.
24.
Campos C, Lajara R, DeLuzio T: Usability and preference assessment of a new prefilled insulin pen versus vial and syringe in people with diabetes, physicians and nurses. Expert Opin Pharmacother 2012;13:1837–1846.
25.
Lajara R, Guerrero G, Thurman J: Healthcare professional and patient perceptions of a new prefilled insulin pen versus vial and syringe. Expert Opin Drug Deliv 2012;9:1181–1196.
26.
Nadeau DA, Campos C, Niemeyer M, et al.: Healthcare professional and patient assessment of a new prefilled insulin pen versus two widely available prefilled insulin pens for ease of use, teaching and learning. Curr Med Res Opin 2012;28:3–13.
27.
Pfützner A, Forst T, Niemeyer M, et al.: Assessment for ease of use and preference of a new prefilled insulin pen (FlexTouch Degludec U100/U200) versus the SoloSTAR insulin pen by patients with diabetes and healthcare professionals. Expert Opin Drug Deliv 2014;11:1381–1389.
28.
Garg S, Bailey T, DeLuzio T, et al.: Preference for a new prefilled insulin pen compared with the original pen. Curr Med Res Opin 2011;27:2323–2333.

Information & Authors

Information

Published In

cover image Diabetes Technology & Therapeutics
Diabetes Technology & Therapeutics
Volume 19Issue Number 10October 2017
Pages: 603 - 607
PubMed: 28704062

History

Published in print: October 2017
Published online: 1 October 2017
Published ahead of print: 13 July 2017

Permissions

Request permissions for this article.

Topics

Authors

Affiliations

Niels Gudiksen, BSc
Device Development, Novo Nordisk A/S, Hillerød, Denmark.
Thibaud Hofstätter, MSc
Device Development, Novo Nordisk A/S, Hillerød, Denmark.
Birgitte B. Rønn, PhD
Biostatistics, Novo Nordisk A/S, Søborg, Denmark.
Thomas Sparre, MD, PhD
Global Development, Medical & Science, Devices & Titration, Novo Nordisk A/S, Søborg, Denmark.

Notes

Address correspondence to:Thomas Sparre, MD, PhDNovo Nordisk A/SVandtårnsvej 1142860 SøborgDenmark
E-mail: [email protected]

Author Disclosure Statement

B. R., N.G., T.H, T.R., and T.S. are employees of Novo Nordisk. T.S and T.H. are shareholders in Novo Nordisk.

Metrics & Citations

Metrics

Citations

Export citation

Select the format you want to export the citations of this publication.

View Options

View options

PDF/EPUB

View PDF/ePub

Get Access

Access content

To read the fulltext, please use one of the options below to sign in or purchase access.

Society Access

If you are a member of a society that has access to this content please log in via your society website and then return to this publication.

Restore your content access

Enter your email address to restore your content access:

Note: This functionality works only for purchases done as a guest. If you already have an account, log in to access the content to which you are entitled.

Media

Figures

Other

Tables

Share

Share

Copy the content Link

Share on social media

Back to Top